| System Function |
| Throughput: | Up to 360 tests per hour |
| Contents of the testing: | Clinical chemistry, Immunoturbidimetric, Electrolyte |
| Analytical modes: | End-point, Two-point, Kinetic, Discrepant Value, Electrolyte (optional) |
| Sample volume: | 4-50ul |
| Reagent volume: | 250-450ul |
| Reaction liquid volume: | 250-500ul |
| Reacting temperature: | 37 ± 0.1°C |
| Programming: | Setting one by one or batching setting. |
| STAT sample priority, Optional ISE unit and barcode reading |
| Sample Tray |
| Sample position: | 118 sample positions (59 sample cup/59 primary tube) |
Support 7ml and 10ml primary tube Any position for standard and quality control setting |
| Reagent Tray |
| Reagent bottle volume: | Reagent one volume 38ml, reagent two volume 18ml.Mono-reagent 60ml |
| Reagent position: | 70 positions (35 dual reagent positions) |
| Reagent cooling temperature: | 5-10°C, 24 hours operating |
| Reagent distribution: | Progress by high precision Syringe, 1ul step |
| Reaction Tray |
| Reaction tray: | 60 cuvettes |
| Reacting temperature: | 37 ± 0.1°C |
| Cuvette cleaning: | Automatic 10 steps cleaning |
| Waste bottle with liquid level sensor |
| Probe |
With preheating function With liquid level sensor, vertical and lateral position sensor |
| Measuring and Optics System |
| Lamp: | Halogen-tungsten lamp |
| Wavelengths: | 8 filters with wavelength from 340~700nm |
| Linearity range: | 0-2.5ABS |
| Resolution: | 0.001ABS |
| User Interface |
User friendly software based on Windows XP operation system Reaction curve, calibration curve and quality control display |